- Product Details
- {{item.text}}
Quick Details
-
Model Number:
-
SK6000
-
Place of Origin:
-
Guangdong, China
-
Reagent tray:
-
30 positions
-
sample tray:
-
60 positions
-
reagent/ sample probe:
-
liquid level detection, collision protection and inventory checking
-
Sample capacity:
-
2-30uL, 0.1ul increasing
-
reagent capacity:
-
30-300ul, 0.1ul increasing
-
reagent proble:
-
sharing with the sample probe
-
Operation system:
-
English operating system, windows
-
test speed:
-
180T/H (without ISE)
-
working mode:
-
discrete/random access
-
minimum reaction volume:
-
150ul
Quick Details
-
Warranty:
-
2 years
-
After-sale Service:
-
Onsite Installation
-
Brand Name:
-
Sinothinker
-
Model Number:
-
SK6000
-
Place of Origin:
-
Guangdong, China
-
Reagent tray:
-
30 positions
-
sample tray:
-
60 positions
-
reagent/ sample probe:
-
liquid level detection, collision protection and inventory checking
-
Sample capacity:
-
2-30uL, 0.1ul increasing
-
reagent capacity:
-
30-300ul, 0.1ul increasing
-
reagent proble:
-
sharing with the sample probe
-
Operation system:
-
English operating system, windows
-
test speed:
-
180T/H (without ISE)
-
working mode:
-
discrete/random access
-
minimum reaction volume:
-
150ul
Mindray chemistry analyzer price / biochemistry analyzer / fully automatic lab equipment
Product Description
Feature:
Open system . You can use any brand's reagents.
Open system . You can use any brand's reagents.
The light source is from the
OSRAM brand
, the filters are a
USA manufacturer
.
Discrete, random access, full automated.
Up to constant 180 tests per hour.
Optional for external bar code reader.
30 positions for samples and 60 positions for reagents.
Automatic probe cleaning, liquid level detection, collision protection.
Reversed optic system with 8 wavelengths(340,405,450,510,546,578,630,670nm) .
Refrigerated reagent and sample compartment .
Operating system:
English operating system: Windows 2000, windows XP, windows7, windows8, windows10 .
Make print format and content freely, provide several common formals.
Special user management password, rational distribution of user operation permissions.
RS-232 standard interface, support for intranet, remote print report.
Optional various databasses, autosave, automatic backup, save all kinds of data permanently.
Real-time monitoring of sample tray, reagent tray, reacting tray; real-time display of reaction temperature, reagent allowance,
reaction curve, calibration curve and quality control chart.
Check and judge reaction endpoint, reaction linear interval, substrate exhaustion, reagent blank absorbance, etc.
Reagents/ sample handing :
Reagent/sample tray: 30 positions for reagents and 60 positions for samples in refrigerated compartment (4~16℃).
Reagent/sample probe: liquid level detection, collision protection and inventory checking.
Probe cleaning: automatic washing for both interior and exterior carry-over<0.1%
Reaction cuvettes cleaning system :
Using unique high efficiency water cleaning method, low carry-over.
Sampling and mixing system :
Sample capacity: 2-30ul, 0.1ul increasing .
Sample probe: sharing with the reagent probe, the inner and auter walls are highly polished, with the function of liquid level
detection, tracking and collision protection, sample probe matching special cleaning liquid, prevent cross contamination.
Reagent capacity: 20-300ul, 0.1ul creasing .
Reagent probe: sharing with the sample probe.
Reagent bottle: reagent bottles with volume of 15ml, 20ml, 30ml, 50ml, 60ml, 70ml can be used, the dead volume is less than 1 ml .
Mixing bar: independent mixing bar, special material, surface treatment using teflon, not hanging liquid, thoroughly clean before
and after mixing to prevent cross contamination.
Calibration and quality control :
Quality control type: real time quality control, daily quality control, day to day quality control .
Quality control charts L-J, cumulative, twin plot, the quality control rules are arbitrarily formulated by default to westguard
multiple rules.
The calibration type: linear and nonlinear. Logit-4P, Logit-5P, Spline, exponent, polynomial, factor method automatically check
the calibration curve and select the best calibration type of the fitting degree automati-cally.
Up to constant 180 tests per hour.
Optional for external bar code reader.
30 positions for samples and 60 positions for reagents.
Automatic probe cleaning, liquid level detection, collision protection.
Reversed optic system with 8 wavelengths(340,405,450,510,546,578,630,670nm) .
Refrigerated reagent and sample compartment .
Operating system:
English operating system: Windows 2000, windows XP, windows7, windows8, windows10 .
Make print format and content freely, provide several common formals.
Special user management password, rational distribution of user operation permissions.
RS-232 standard interface, support for intranet, remote print report.
Optional various databasses, autosave, automatic backup, save all kinds of data permanently.
Real-time monitoring of sample tray, reagent tray, reacting tray; real-time display of reaction temperature, reagent allowance,
reaction curve, calibration curve and quality control chart.
Check and judge reaction endpoint, reaction linear interval, substrate exhaustion, reagent blank absorbance, etc.
Reagents/ sample handing :
Reagent/sample tray: 30 positions for reagents and 60 positions for samples in refrigerated compartment (4~16℃).
Reagent/sample probe: liquid level detection, collision protection and inventory checking.
Probe cleaning: automatic washing for both interior and exterior carry-over<0.1%
Reaction cuvettes cleaning system :
Using unique high efficiency water cleaning method, low carry-over.
Sampling and mixing system :
Sample capacity: 2-30ul, 0.1ul increasing .
Sample probe: sharing with the reagent probe, the inner and auter walls are highly polished, with the function of liquid level
detection, tracking and collision protection, sample probe matching special cleaning liquid, prevent cross contamination.
Reagent capacity: 20-300ul, 0.1ul creasing .
Reagent probe: sharing with the sample probe.
Reagent bottle: reagent bottles with volume of 15ml, 20ml, 30ml, 50ml, 60ml, 70ml can be used, the dead volume is less than 1 ml .
Mixing bar: independent mixing bar, special material, surface treatment using teflon, not hanging liquid, thoroughly clean before
and after mixing to prevent cross contamination.
Calibration and quality control :
Quality control type: real time quality control, daily quality control, day to day quality control .
Quality control charts L-J, cumulative, twin plot, the quality control rules are arbitrarily formulated by default to westguard
multiple rules.
The calibration type: linear and nonlinear. Logit-4P, Logit-5P, Spline, exponent, polynomial, factor method automatically check
the calibration curve and select the best calibration type of the fitting degree automati-cally.
working mode
|
discrete/random access
|
test speed
|
test speed
180T/H
(without ISE)
|
Test principle
|
absorbance photometry, turbidimetry
|
Methodology
|
end-point, fixed-time, kinetic, single/dual reagent chemistries, monochromatic/ bichromatic linear/non-linear, multi-point
calibration |
the longest reaction time
|
15 minutes
|
minumum reaction volume
|
150ul
|
cuvettes material
|
plastics (quartz glass can be selected),
81 cuvettes
have the function of automatically deducing.
|
reagent and sample blanks linear range of absorbance
|
0-3.8Abs, halogen light source using time≥
2000 hours
|
Wavelength
|
340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 670nm
|
Refrigeration system
|
water medium uniform refrigeration technology
|
Reagent storehouse temperature
|
4-16℃
|
Reaction temperature
|
37±0.2℃
|
Water consumption
|
5L/hour
|
Power supply
|
200~240V 50/60HZ 600W
|
Dimension
|
834mm * 643mm * 756mm
|
Rough weight
|
91kg
|
Net weight
|
65kg
|
Temperature fluctuation
|
±0.1℃
|