- Product Details
- {{item.text}}
Quick Details
-
Model Number:
-
UM-5
-
Product name:
-
Portable Ultrasonic Thickness Gauge
-
Display Type:
-
2.4 color OLED 320*240 pixels, contrast 10000:1
-
Operating Principle:
-
Pulse echo with dual element transducers
-
Unit:
-
Mm/inch (selectable)
-
Rectify Modes:
-
RF+, RF-, HALF+, HALF-, FULL
-
Measuring range:
-
0.5mm to 508mm (0.02'' to 20.00'')
-
Update Rate:
-
Selectable 4Hz, 8Hz, 16Hz
-
Resolution:
-
0.01mm, 0.1mm
-
Size:
-
153mm*76mm*37mm
-
Weight:
-
280g including batteries
Quick Details
-
Warranty:
-
1 year
-
Place of Origin:
-
China
-
Brand Name:
-
YUSHI
-
Model Number:
-
UM-5
-
Product name:
-
Portable Ultrasonic Thickness Gauge
-
Display Type:
-
2.4 color OLED 320*240 pixels, contrast 10000:1
-
Operating Principle:
-
Pulse echo with dual element transducers
-
Unit:
-
Mm/inch (selectable)
-
Rectify Modes:
-
RF+, RF-, HALF+, HALF-, FULL
-
Measuring range:
-
0.5mm to 508mm (0.02'' to 20.00'')
-
Update Rate:
-
Selectable 4Hz, 8Hz, 16Hz
-
Resolution:
-
0.01mm, 0.1mm
-
Size:
-
153mm*76mm*37mm
-
Weight:
-
280g including batteries
1. Color OLED 320x240 Pixels Display
2. Live Simultaneous Color A-Scan and Thickness Display
3. Control of Gain, Blanking, Gate, Range, Delay, RF and Rectify Modes
4. Time-based B-Scan Display
5. Echo-Echo Mode(Thru Paint & Coatings)
6. Stores 100,000 Readings and 1000 waveform
Live Color A-Scan
Users could directly see the color waveform of the ultrasonic sound (or A-scan) on the screen, which is quite important for the occasions that we need to check the correctness of the testing results. Many cases will cause wrong testing results or even noreadings. We could find the causes easily through the A-scan. Adjust the three parameters of GAIN, BLACKING, GATE, and then we will get the right readings.
Live Color B-Scan
UM-5 thickness gauge has time base B-scan function. Move the probe along the workpiece surface, then the cross -sectional profile of the workpiece display, use for observe the underside contour of the workpiece.It could be automatically capture a minimum value of the B-scan image, and indicate the position of the minimum by a red triangle. You can see any point thickness value of the B-scan image by moving the pointer.
More Practical Functions:
BLANKING;GATE; RED ARROW; RANGE; DELAY; RecitifIcation Modes:four modes selective; DIFFERENCE/RATE DIFF/RR%; Max./Min. Capture; Alarm Mode; Update Rate; Multi- Languages Available etc.
Display Type | 2.4'' color OLED, 320×240 pixels, contrast 10,000:1 |
Operating Principle | Pulse echo with dual elemet transducers |
Measuring Range | 0.50mm to 508mm(0.02'' to 20.00''), depending on material, probe and surface condition |
Measuring Resolution | Selectable 0.01mm, 0.1mm(selectable 0.001'', 0.01'') |
Units | Inch or Millimeter |
Rectify Mode | RF+, RF-, HALF+, HALF-, FULL |
Display Mode | Normal, Minimum/Maximum capture, DIFF/RR%, A-Scan, B-Scan |
V-Path Correction | Automatic |
Update Rate | Selectable 4Hz, 8Hz, 16Hz |
Material Velocity Range | 500 to 9999m/s(0.0197 to 0.3939 in/μs) |
Languages | English, French, German, etc. |
Alarm Settings | Minimum and Maximum alarms. Range of 0.25mm to 508 mm(0.010'' to 20.00''). Dynamic waveform color change on alarm |
Power Requirements | 2 AA size batteries |
Operating Time | Approximately 40 hours |
Instrument Shut-off | Selectable ALWAYS ON or AUTO OFF after 5, 10, 20 minutes of inactivity |
Operating Temperature | -10℃ to +50℃(+10°F to +12°F ) |
Size | 153mm × 76mm × 37mm (H × W × D) |
Weight | 280g including batteries |
Certification
Standard Configuration of UM-5 Ultrasonic Thickness Gauge
Name | Quantity |
UM-5 Ultrasonic Thickness Gauge | 1 |
Probe | 1 |
Battery | 2 |
Couplant | 1 |
Carrying Case | 1 |
Operating Manual | 1 |
USB Cable | 1 |
Software | 1 |