- Product Details
- {{item.text}}
Quick Details
-
Brand Name:
-
FORERMED
-
Model Number:
-
YJ-E800B
-
Product name:
-
Electrosurgical unit (ESU)
-
Screen:
-
Touch screen
-
Input Voltage:
-
100-240V
-
Input Frequency:
-
50/60Hz
-
Working Frequency:
-
512KHz, 300KHz
-
bipolar coagulation:
-
80W
-
spray coagulation:
-
100W
-
forced coagulation:
-
120W
-
soft coagulation:
-
150W
-
hyper cut:
-
300W
Quick Details
-
Warranty:
-
1 Year
-
After-sale Service:
-
Online technical support
-
Place of Origin:
-
Henan, China
-
Brand Name:
-
FORERMED
-
Model Number:
-
YJ-E800B
-
Product name:
-
Electrosurgical unit (ESU)
-
Screen:
-
Touch screen
-
Input Voltage:
-
100-240V
-
Input Frequency:
-
50/60Hz
-
Working Frequency:
-
512KHz, 300KHz
-
bipolar coagulation:
-
80W
-
spray coagulation:
-
100W
-
forced coagulation:
-
120W
-
soft coagulation:
-
150W
-
hyper cut:
-
300W
High frequency electrosurgical system is……?
A kind of electrosurgical instrument (patent number CN201 620756399.0) to replace the mechanical
scalpel for tissue cutting.
It heats up the tissue by contacting the body with the high-frequency and high-voltage current generated by
the tip of the effective electrode to achieve the separation and coagulation of the body tissue, so as to
achieve the purpose of cutting and hemostasis.
scalpel for tissue cutting.
It heats up the tissue by contacting the body with the high-frequency and high-voltage current generated by
the tip of the effective electrode to achieve the separation and coagulation of the body tissue, so as to
achieve the purpose of cutting and hemostasis.
Product Description
High-frequency Electrosurgical Unit YJ-E800B
Feature
* Automatic control technology, accurate and fast automatic adjustment of parameters
* Better operation visual field and less smell during operation
* People oriented low-risk design, various monitoring systems to reduce the risk of surgery
* Better operation visual field and less smell during operation
* People oriented low-risk design, various monitoring systems to reduce the risk of surgery
Applicable department
General surgery, orthopedics, obstetrics and gynecology, urology, cardiothoracic surgery, anorectal surgery
Available in Neurosurgery, microsurgery, ENT, plastic surgery and etc. under low power mode.
Cooperate with endoscope, laparoscope, cystoscope for minimally invasive and intracavitary surgery
Available in Neurosurgery, microsurgery, ENT, plastic surgery and etc. under low power mode.
Cooperate with endoscope, laparoscope, cystoscope for minimally invasive and intracavitary surgery
Technical Specifications
Technical parameters
|
|
|
Input Voltage
|
100-240V
|
|
Input Frequency
|
50/60Hz
|
|
Working Frequency
|
512KHz,300KHz
|
|
Working Mode
|
High cut, pure cut, blended cut, endoscopically cut, soft coagulated, forced coagulated, jet coagulated, bipolar coagulated
|
Working mode
|
Parameter
|
|
|
hyper cut
|
300W
|
400Ω
|
|
pure cut
|
250W
|
400Ω
|
|
mixed cut 1
|
150W
|
400Ω
|
|
mixed cut 2
|
120W
|
400Ω
|
|
mixed cut 3
|
80W
|
400Ω
|
|
soft coagulation
|
150W
|
400Ω
|
|
forced coagulation
|
120W
|
400Ω
|
|
spray coagulation
|
100W
|
400Ω
|
|
bipolar coagulation
|
80W
|
50Ω
|
High frequency electric knife working modes
|
• Intelligent Cutting
|
High-temperature resection:
This mode can be used in underwater vaporization resection to achieve ideal cutting effect
|
Pure resection:
Similar to the real scalpel cutting, it is an accurate non-destructive cutting for any tissues
|
Mixed resection:
Cutting with coagulation process
|
* Intelligent Coagulation
|
|
|
Forced Coagulation
Point coagulation - contact hemostasis, efficient coagulation and electrocoagulation result |
Spray Coagulation
Area coagulation - The surgical electrode no need to touch biological tissues, it is suitable for hemostasis on a large surface (it can produce argon ions for electrocoagulation by cooperationg with an argon controller) |
|
Bipolar Coagulation
For nerosurgical or general bipolar electrocoagulation procedures, achieving tissue or vascular coagulation between two forceps |
Soft Coagulation
Rapid contact coagulation - good for the healing after surgery, it is suitable for endoscopic surgery and other accurate tissue rapid contact coagulation |
Components for high frequency electrosurgical systems
The electrode will provide intimate applicability and conveniency, lower risk, longer service life, and excellent electrocoagulation and cutting effect
Details Images