- Product Details
- {{item.text}}
Quick Details
-
Resolution:
-
0.1nT
-
Accuracy:
-
< 0.5% of reading
-
Size:
-
320*130*285
-
Axis:
-
3
-
Power supply:
-
220V AC
-
Bandwidth:
-
>1.5kHz or >100Hz
-
Certification:
-
IECEE
Quick Details
-
Warranty:
-
1 years
-
Place of Origin:
-
Fujian, China
-
Brand Name:
-
Dexing
-
Resolution:
-
0.1nT
-
Accuracy:
-
< 0.5% of reading
-
Size:
-
320*130*285
-
Axis:
-
3
-
Power supply:
-
220V AC
-
Bandwidth:
-
>1.5kHz or >100Hz
-
Certification:
-
IECEE
DX-330F High Resolution Three-Axis Fluxgate Magnetometer
Dexing flux - gate magnetometers are instruments for accurate measurement of weak static and dynamic and low frequency vector magnetic fields. Compared with a magnetic field measuring instrument with high stability, linearity and accuracy, it is compared with the principle of Hall effect or the principle of magneto-resistance effect. It is the best choice to measure the weak magnetic field , with the output of digital and analog signals , the unique CAN bus network interface is suitable for multi - array magnetic field testing , and is widely used in the fields of scientific research , military industry , aerospace , etc .
Hardware interface information of Flux-gate Magnetometer
Interface DB15 |
Probe input/output |
Measuring mode |
Hand-held probe or homemade tooling |
Reading mode |
Display screen or computer software |
Display |
7-bit SVG color screen |
BNC |
Analog signal output |
Power supply |
220V |
RS-232 |
Digital signal input / output (providing instruction format) |
Parameter
DX-330F |
||||
Size(mm)(width * height * depth) |
320*130*285 |
|||
Dimensions |
3 |
|||
Resolution |
0.1nT |
|||
Preheating Time |
15mins |
|||
Accuracy |
< 0.5% of reading |
|||
Noise level |
10 -20pTrms/√Hz at 1Hz |
|||
Range (optional) |
±100μT |
±500μT |
±1000μT |
|
Error of temperature coefficient |
1nT/ºC |
±10nT/°C |
±10nT/°C |
|
Bandwidth at -3db |
>1.5kHz |
>100Hz |
||
Three-dimensional orthogonality error |
<1° |
<2° |
||
Frequency response (DC deviation: < 5) |
1000Hz |
1000Hz |
||
Operating temperature range |
-40ºC to +70ºC |
0°C to +215°C |
||
Acquisition rate |
0-300 times/second |
0-300 times/second |
||
Probe size(mm) (original British) |
32*32*122 |
23*23*21.4 |
||
Communication |
RS-232/USB |
RS-232/USB |
||
Analog output |
3 channels BNC |
3 channels BNC |
||
Supporting software |
Data reading software |
Data reading software |
B ackpanel
1. Power Interface and Switch
2. Rs-232c Connector
3. X Channel Analog Output BNC
4. Y Channel Analog Output BNC
5. Z Channel Analog Output BNC
6. Probe Input Connector
Applications:
♦ Monitoring the magnetic field of earth environment, laboratory, and space.
♦ Calibrating laboratory magnetic field shielding and generating field sources, such as: 3D Helmholtz coil system, non-distance coil system etc.
♦ Geological prospecting. Petroleum system while drilling system detect weak magnetic field in rock.
♦ Testing the attenuation characteristics and shielding effectiveness of magnetic field shielding system.
♦ Detecting magnetic shielding effect in magnetic shielding room.
♦ Dipole moment and remanence magnetic field of satellite.
♦ Calibration standard for flux-gate meter, magneto-optical effect magnetometer and all kinds of sensors.
♦ Earthquake precursor monitoring, volcanic observation, and other environmental and disaster geological work.
♦ Aeronautical and marine magnetic measurements.
♦ Research on weak magnetic field in bioscience.
Hongkong Exhibition