- Product Details
- {{item.text}}
Quick Details
-
Brand Name:
-
MSL
-
Model Number:
-
MSLDX20
-
Product name:
-
Dental CBCT Machine
-
Tube type:
-
D-054SB
-
Tube voltage:
-
60 - 90kV
-
Maximum tube current:
-
10mA
-
Target angle:
-
5°
-
Focus spot:
-
0.5mm
-
Reference axis:
-
Middle point of the bottom edge in the effective area
-
Total filtration:
-
2.7mm Al
-
Nominal supply voltage:
-
200 - 240V / 90 - 110V
-
Input power frequency:
-
50 / 60 Hz
Quick Details
-
Warranty:
-
1 Year
-
After-sale Service:
-
Online technical support
-
Place of Origin:
-
Guangdong, China
-
Brand Name:
-
MSL
-
Model Number:
-
MSLDX20
-
Product name:
-
Dental CBCT Machine
-
Tube type:
-
D-054SB
-
Tube voltage:
-
60 - 90kV
-
Maximum tube current:
-
10mA
-
Target angle:
-
5°
-
Focus spot:
-
0.5mm
-
Reference axis:
-
Middle point of the bottom edge in the effective area
-
Total filtration:
-
2.7mm Al
-
Nominal supply voltage:
-
200 - 240V / 90 - 110V
-
Input power frequency:
-
50 / 60 Hz
Multifunctional Dental CBCT Equipment
MSLDX20
Product Description
Technical Parameter
Tube assembly
|
|
|
Tube type
|
D-054SB
|
|
Tube voltage
|
60 - 90kV
|
|
Maximum tube current
|
10mA
|
|
Target angle
|
5°
|
|
Focus spot
|
0.5mm
|
|
Reference axis
|
Middle point of the bottom edge in the effective area
|
|
Total filtration
|
2.7mm Al
|
|
Electrical connections
|
|
|
Nominal supply voltage
|
200 - 240V / 90 - 110V
|
|
Input power frequency
|
50 / 60 Hz
|
|
Nominal current
|
8.2A @ 220V, 18A @ 100V
|
|
Fuses
|
220 Vac:
Hollyland 61T-160H 100 Vac:
Hollyland 65TS-200L
|
|
Power consumption
|
1800VA
|
|
Maximum impedance of
mains
|
0.2Ω
|
|
Positioning laser lamp
|
|
|
Frankfurt plane laser light
Median plane laser light
|
Class 1 laser products
Warning symbols are placed next to the laser light.
|
|
X-ray generator
|
|
|
Generator
|
High frequency DC generator
|
|
Tube voltage
|
60 - 90kV (±5kV)
|
|
Tube current
|
2 - 10mA (±1mA)
|
|
Accuracy of exposure
duration
|
±2%
|
|
Accuracy of
dose-area-product
|
±50%
|
|
Radiation output accuracy
|
> 90%
|
|
Physical measures
|
|
|
Source-image distance
(SID)
|
584mm (PA)
1710mm (CE)
|
|
Source-object distance
(SOD)
|
441mm ( PA)
1285 (CE)
|
|
X-ray beam dimensions
|
PA:
|
Images