- Product Details
- {{item.text}}
Quick Details
-
Place of Origin:
-
Shaanxi, China
-
Product name:
-
Micro Pipette
-
Volume Range:
-
0.1-2.5μl
-
Material:
-
Plastic
-
Type:
-
Adjustable Type
-
Application:
-
Medical, Laboratory
-
Certificate:
-
ISO
-
Color:
-
White and gray
-
Increment:
-
0.05μL
-
Test Volume:
-
2.5μL
-
Maximum permissible systematic error:
-
0.0625μL
Quick Details
-
Warranty:
-
1 year
-
Brand Name:
-
TOPTION
-
Model Number:
-
TOPT-7010101001
-
Place of Origin:
-
Shaanxi, China
-
Product name:
-
Micro Pipette
-
Volume Range:
-
0.1-2.5μl
-
Material:
-
Plastic
-
Type:
-
Adjustable Type
-
Application:
-
Medical, Laboratory
-
Certificate:
-
ISO
-
Color:
-
White and gray
-
Increment:
-
0.05μL
-
Test Volume:
-
2.5μL
-
Maximum permissible systematic error:
-
0.0625μL
New cal pipette filler dropper for lab
Description of the New cal pipette filler dropper for lab
♦ This Micro Pipette is Single-channel Adjustable Micro Pipette,it have been quality tested according to ISO8655-2:2002 with calibration certificate.The quality control involves gravimetric testing of each pipette with distilled water at 22℃.It's widely used in the clinical diagnostic laboratories, biotechnology laboratories, aceutical and chemical laboratories, environmental laboratories and food laboratories.
♦ Except the Single-channel Adjustable Micro Pipette, we also supply single-channel fixed pipettes,multichannel pipettes.
Features of this New cal pipette filler dropper for lab
♦ Lightweight, ergonomic, low force design
♦ Digital display clearly reads volume setting
♦ The pipettes cover volume range of 0.1μl to 10ml
♦ Easy to calibrate and maintain with tool supplied
♦ Design helps avoid repetitive strain injuries
♦ Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate.
♦ The low part is available for autoclaving
Specificaation of this New cal pipette filler dropper for lab
Single-channel Adjustable Volume Pipettes | ||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | ||
% | μL | % | μL | |||
0.1-2.5μL | 0.05μL |
2.5μL |
2.50% |
0.0625 |
2.00% |
0.05 |
1.25μL |
3.00% |
0.0375 |
3.00% |
0.0375 |
||
0.25μL |
12.00% |
0.03 |
6.00% |
0.015 |
||
0.5-10μL |
0.1μL |
10μL |
1.00% |
0.1 |
0.80% |
0.08 |
5μL |
1.50% |
0.075 |
1.50% |
0.075 |
||
1μL |
2.50% |
0.025 |
1.50% |
0.015 |
||
2-20μL |
0.5μL |
20μL |
0.90% |
0.18 |
0.40% |
0.08 |
10μL |
1.20% |
0.12 |
1.00% |
0.1 |
||
2μL |
3.00% |
0.06 |
2.00% |
0.04 |
||
5-50μL |
0.5μL |
50μL |
0.60% |
0.3 |
0.30% |
0.15 |
25μL |
0.90% |
0.225 |
0.60% |
0.15 |
||
5μL |
2.00% |
0.1 |
2.00% |
0.1 |
||
10-100μL |
1μL |
100μL |
0.80% |
0.8 |
0.15% |
0.15 |
50μL |
1.00% |
0.5 |
0.40% |
0.2 |
||
10μL |
3.00% |
0.3 |
1.50% |
0.15 |
Shipping of the New cal pipette filler dropper for lab
♦ We can ship the goods to you by International Express such as DHL,UPS,TNT,EMS and so on, you can also choose the appropriate one based on your timeline and budget. Besides,you can choose to use your own forwarder.