Brand Name:
-
OTS
-
Model Number:
-
XB-OTS-120
-
Product name:
-
salt spray test machine / spray nozzle / salt spray test chamber
-
spray volume::
-
1.0~2.0ml/80cm2/h
-
Laboratory Temperature::
-
NSS.ACSS 35°C ±1°C/ CASS 50°C ±1°C
-
PH::
-
6.5~7.2 3.0~3.2
-
air pressure::
-
86~ 106Kpa
-
Temp. air::
-
NSS.ACSS 47°C±1°C/CASS 63°C±1°C
-
Standards::
-
ASTM B117, GB/T2423.17
-
Uses::
-
Electronic Salt Fog Testing Cabinet
-
Power::
-
Electronic, AC220V 50HZ/60HZ 4A
-
Spray method::
-
Continuous spraying
Quick Details
-
Warranty:
-
14 months
-
Power:
-
Electronic
-
Place of Origin:
-
Guangdong, China
-
Brand Name:
-
OTS
-
Model Number:
-
XB-OTS-120
-
Product name:
-
salt spray test machine / spray nozzle / salt spray test chamber
-
spray volume::
-
1.0~2.0ml/80cm2/h
-
Laboratory Temperature::
-
NSS.ACSS 35°C ±1°C/ CASS 50°C ±1°C
-
PH::
-
6.5~7.2 3.0~3.2
-
air pressure::
-
86~ 106Kpa
-
Temp. air::
-
NSS.ACSS 47°C±1°C/CASS 63°C±1°C
-
Standards::
-
ASTM B117, GB/T2423.17
-
Uses::
-
Electronic Salt Fog Testing Cabinet
-
Power::
-
Electronic, AC220V 50HZ/60HZ 4A
-
Spray method::
-
Continuous spraying
The salt spray test chamber
is a climate simulation device that tests the corrosion of electronic equipment, metallic material, as well as some surface treatment--coating, galvanizing, organic/inorganic coating, anodizing treatment, rust-proof oil coating, etc. Performance indicators in line with requirements of GB / T5170.8 Electrical and Electronic Products, Environmental Testing Equipment, the Basic Parameters of the Test Method. The salt spray corrosion test chamber according to GB / T2423.17 Electronic and Electrical Products Basic Environmental Test Procedures Test: Salt Spray Test Method for neutral salt spray test can also be used for acetic acid salt test.
Model
|
XB-OTS-60
|
XB-OTS-90
|
XB-OTS-120
|
XB-OTS-160
|
XB-OTS-200
|
||||
Inner size (cm)
|
60x45x40
|
90x60x50
|
120x100x50
|
160x100x50
|
200x100x50
|
||||
External size(cm)
|
107x60x118
|
141x88x128
|
190x110x140
|
230x130x140
|
270x150x150
|
||||
Inside temp
|
Brine test method(NSS ACSS)35°C±1°C/Corrosion resistance test method
(CASS)50°C±1°C |
|
|
|
|
||||
Brine temp
|
(NSS ACSS) 35°C±1°C (CASS) 50°C±1°C
|
|
|
|
|
||||
Test chamber volume
|
108L
|
270L
|
600L
|
800L
|
1000L
|
||||
Brine chamber volume
|
15L
|
25L
|
40L
|
40L
|
40L
|
||||
Compressed air pressure
|
1.00±0.01kgf/cm2
|
|
|
|
|
||||
Spray volume
|
1.0~2.0ml/80cm2/h (Collect at least 16 hours, the average)
|
|
|
|
|
||||
Test room relative humidity
|
85%RH or more
|
|
|
|
|
||||
PH
|
6.5~7.2 3.0~3.2
|
|
|
|
|
||||
Spray method
|
Programmable spray (including the continuous spray interrupted spray)
|
|
|
|
|
||||
Power
|
AC220V1Φ10A
|
AC220V1Φ15A
|
AC220V1Φ20A
|
AC220V1Φ20A
|
AC220V1Φ30A
|
Advantage
Ø 24 hours online
Ø 35 years experiences
Ø Sample chamber of each model
Ø Best price, fast delivery
Ø OEM,ODM
Ø Oversea offices
Ø Own import and export business license
Ø ISO,CE,UL,ASTM,DIN,EN,GB,BS,JIS,ANSI,TAPPI,AATCC,IEC,VDE
Ø 35 years experiences
Ø Sample chamber of each model
Ø Best price, fast delivery
Ø OEM,ODM
Ø Oversea offices
Ø Own import and export business license
Ø ISO,CE,UL,ASTM,DIN,EN,GB,BS,JIS,ANSI,TAPPI,AATCC,IEC,VDE
Customer Support Services
Ø Installation
Ø Training (Training customer employees)
Ø Calibration
Ø Preventative maintenance
Ø Replacement parts
Ø Assistance via phone or internet
Ø On-site diagnosis and repair/online diagnosis and repair
Ø Training (Training customer employees)
Ø Calibration
Ø Preventative maintenance
Ø Replacement parts
Ø Assistance via phone or internet
Ø On-site diagnosis and repair/online diagnosis and repair
Recommend Products
Hot Searches