- Product Details
- {{item.text}}
Quick Details
-
Brand Name:
-
Huazheng
-
Model Number:
-
HZJD8001
-
name:
-
Electrical Multimeter and Clamp Meter Calibrator
Quick Details
-
Warranty:
-
1 Years
-
Power:
-
Electronic
-
Place of Origin:
-
Hebei, China
-
Brand Name:
-
Huazheng
-
Model Number:
-
HZJD8001
-
name:
-
Electrical Multimeter and Clamp Meter Calibrator
Huazheng Electric HZJD8001 Multimeter And Clamp Meter Calibrator Device for Verify Various Ammeters, Voltmeter, Resistance Meter
I . Use and characteristics
HZJD8001 multi-function calibrator is suitable for verifying and calibrating various ammeters, voltmeters and resistance meters below 0.2 level. The JJF 1587-2016 digital multimeter calibration specification can also be used as a high-stability test power supply. It can be used with a high-level standard meter to proofread ammeters, voltmeters, and resistance meters above 0.2. This calibrator has two amplitude adjustment methods: analog adjustment (potentiometer) and digital adjustment (keyboard). The clamp meter can also be calibrated, in line with the JJF 1075-2015 clamp-type ammeter calibration specification.
II. Main functions
1. 5 1 /2 -digit LED digital display output, which can be displayed in two ways: actual value or percentage.
2. The output range of AC and DC current is: 0~100uA~500uA~2mA~5mA~20mA~50mA~200mA~500mA~2A~5A~20A.
3. The output range of AC and DC voltage is:
0~200mV~1V~2V~5V~10V~20V~50V~100V~200V~500V~1000V.
There are three kinds of AC frequencies for selection: 50 Hz, 60 Hz and 400 Hz.
The output overload can be automatically protected and reset manually.
6. Each voltage and current can be overloaded by 10%.
7. Clamp meter measurement: Equipped with our standard coil, it can measure AC 0~750A (some meters can measure above 1000A); DC 0~1000A, with an error of ±0.3%.
III.Main technical indicators
1. Stability: DC 0.01% of full capacity / 2 minutes
AC 0.03% of full capacity / 2 minutes
2.AC Distortion: 0.5%
3. DC ripple factor: 0.1%
4. Output frequency and accuracy: 50 Hz, 60 Hz, 400 Hz 1%
5.The range and accuracy of output voltage and current (23 ℃± 2 ℃ , the output value is greater than 10% of the range).
6. Power consumption: AC power voltage 220V ± 10%, frequency 50 Hz ± 1 Hz; power 180VA.
7.Working environment: The temperature of the working environment is 5 ℃ ~35 ℃ , and the relative humidity is not more than 80%.
8. Working hours: Continuous.
9. Dimensions: 135 × 480 × 420mm3
10. Weight: about 16kg.
Output item |
Output range |
Rated output |
Max output |
Accuracy |
AC/DC voltage |
0~200mV |
10mA |
30mA |
DC: ±(0.05% of reading + 0.02% of full scale) ±0.02mV AC: ±(0.07% of reading + 0.03% of full scale) ±0.04mV |
0~2V |
20mA |
60mA |
||
0~5V~10V~20V |
40mA |
100mA |
||
0~200V~500V |
20mA |
60mA |
||
0~1000V |
10mA |
30mA |
||
AC/DC current |
0~100uA ~500uA ~2mA ~5mA ~20mA~50mA |
DC: 3V AC: 36V |
DC: ±(0.07% of reading + 0.03% of full scale) AC: ±(0.07% of reading + 0.03% of full scale) Accuracy under 2mA is not tested |
|
0~200mA ~500mA |
DC: 3V AC: 12V |
|||
0~2A~5A~20A |
2V |
|||
Standard resistance |
10, 24 , 50 , 100 , 240 , 500 ( Ω ) 1, 2.4 , 5 , 10 , 24 ( KΩ ) Plus ×1K, ×1 magnification gear (ie 10KΩ~24MΩ) |
0.25W |
± 0.2%+20mΩ |